Free Triiodothyronine (fT3) ELISA Kit

 
Catalog #
EKCA1019
Species
Human
 
Be the first to write a review
 Catalog #AvailabilitySizeQuantityUnit Price Save For Later Wish List
EKCA1019-1 2 weeks 96 T call
Select product before adding to cart
  • Datasheet
  • Citations
  • Reviews
NameFree Triiodothyronine (fT3) ELISA Kit
INTENDED USEThe Calbiotech, Inc. () fT3 ELISA kit is used for the quantitative measurement of free Triiodothyronine (fT3) in human serum.
SUMMARY AND EXPLANATIONOver 99% of Triiodothyronine (T3) circulates in blood is bound to carrier proteins; thyroxine-binding globulin (TBG). However, only the free (unbound) portion of T3 is responsible for the biological action. Further, the concentrations of the carrier proteins are altered in many clinical conditions, such as pregnancy. In normal thyroid function as the concentrations of the carrier proteins alters, the total T3 level changes so that the freeT3 concentration remains constant. Thus, measurements of free T3 concentrations correlate more reliably with clinical status than total T3 levels. The increase in total T3 levels associated with pregnancy, oral contraceptives and estrogen therapy result in higher total T3 levels while the free T3 concentration remains basically unchanged.
PRINCIPLE OF THE TESTThe fT3 is a solid phase competitive ELISA. The samples Anti-T3 Biotin and fT3 enzyme conjugate are added to the wells coated with Streptavidin. fT3 in the patient’s serum competes with a T3 enzyme conjugate for binding sites. Unbound T3 and T3 enzyme conjugate is washed off by washing buffer. Upon the addition of the substrate, the intensity of color is inversely proportional to the concentration of fT3 in the samples. A standard curve is prepared relating color intensity to the concentration of the fT3.
MATERIALS PROVIDED 96 Tests
Microwells coated with Streptavidin 12x8x1
fT3 Standard: 6 vials ( ready to use) 0.5 ml
Anti-T3 Biotin Solution 7 ml
fT3 Enzyme conjugate: 1 Bottle (ready to use) 7 ml
Substrate: 1 bottle (ready to use) 12 ml
Stop Solution: 1 bottle (ready to use) 12 ml
20X Wash concentrate: 1 bottle 25 ml
MATERIALS NOT PROVIDEDDistilled or deionized water
Precision pipettes
Disposable pipette tips
ELISA reader capable of reading absorbance at 450nm
Absorbance paper or paper towel
Graph paper
STORAGE AND STABILITY1. Store the kit at 2 – 8 C.
Keep microwells sealed in a dry bag with desiccants.
The reagents are stable until expiration of the kit.
Do not expose test reagents to heat, sun, or strong light.
WARNINGS PRECAUTIONS
1. Potential biohazardous materials:
The calibrator and controls contain human source components, which have been tested and found non-reactive for hepatitis B surface antigen as well as HIV antibody with FDA licensed reagents. However there is no test method that can offer complete assurance that HIV, Hepatitis B virus or other infectious agents are absent. These reagents should be handled at the Biosafety Level 2, as recommended in the Centers for Disease Control/National Institutes of Health manual, "Biosafety in Microbiological and Biomedical Laboratories" 1984.
2. This test kit is USA FDA exempt product.
3. Do not pipette by mouth. Do not smoke, eat, or drink in the areas in which specimens or kit reagents are handled.
The components in this kit are intended for use as an integral unit. The components of different lots should not be mixed.
It is recommended that standards, control and serum samples be run in duplicate.
Optimal results will be obtained by strict adherence to this protocol. Accurate and precise pipetting, as well as following the exact time and temperature requirements prescribed are essential. Any deviation from this may yield invalid data.
SPECIMEN COLLECTION HANDLINGCollect blood specimens and separate the serum immediately.
Specimens may be stored refrigerated at (2-8 C) for 5 days. If storage time exceeds 5 days, store frozen at (-20 C) for up to one month.
Avoid multiple freeze-thaw cycles.
Prior to assay, frozen sera should be completely thawed and mixed well.
Do not use grossly lipemic specimens.
REAGENT PREPARATIONPrepare 1X Wash buffer by adding the contents of the bottle (25 ml, 20X) to 475 ml of distilled or deionized water. Store at room temperature.
ASSAY PROCEDUREBring all specimens and kit reagents to room temperature (20-25C) and gently mix.
Format the microplates wells for control, standard and patient samples to be assayed in duplicate. Place any unused microwell strips back into the aluminum bag, seal and store at 2-8°C.
Pipette 50 l of fT3 standards, control and samples into the assigned well.
Add 50 l of fT3 enzyme conjugate to all wells.
Add 50 l of Anti-T3 Biotin Solution to all the wells.
Incubate for 60 minutes at room temperature (20-25C).
Remove liquid from all wells. Fill wells with 300 l 1X wash buffer (see buffer preparation above) Wash three times. Blot on absorbent paper towels.
Add 100 l of substrate to all wells.
Incubate for 15 minutes at room temperature.
Add 50 l of stop solution to all wells. Shake the plate gently to mix the solution.
Read absorbance on ELISA Reader at 450 nm within 15 minutes after adding the stopping solution.
CALCULATION OF RESULTSThe standard curve is constructed as follows:
Check fT3 standard value on each standard vial. This value might vary from lot to lot. Make sure you check the value on every kit.
To construct the standard curve, plot the absorbance for fT3 standards (vertical axis) versus fT3 standard concentrations (horizontal axis) on a linear graph paper. Draw the best curve through the points.
Read the absorbance for controls and each unknown sample from the curve. Record the value for each control or unknown sample.
EXAMPLE OF STANDARD CURVE
OD 450 nm Conc. pg/mL
Std 1 2.467 0
Std 2 2.149 1.2
Std 3 1.823 2.5
Std 4 1.383 5.0
Std 5 0.976 8.5
Std 6 0.591 18.0
LIMITATIONS OF THE TESTThe test results obtained using this kit serve only as an aid to diagnosis and should be interpreted in relation to the patient’s history, physical findings and other diagnostic procedures.
Do not use sodium azide as preservative. Sodium azide inhibits HRP enzyme activities.
EXPECTED VALUESIt is recommended that each laboratory establish its own normal ranges based on a representative sampling of the local population. The following values for fT3 were established by the and may be used as initial guideline ranges only:
Classification pg/ml
Adult 1.4-4.2
PERFORMANCE CHARACTERISTICSCorrelation with a Reference ELISA kit:
A total of 128 sera were tested by fT3 ELISA and a reference ELISA kit. Results were as follows:
Correlation Slope Intercept
0.95 0.925 0.15

Precision
Intra-Assay Precision was determined by assaying 16 replicates of each of three sera; normal, low and high.
Serum No. of Replicates Mean
pg/ml
Standard Deviation Coefficient of Variation (%)
1 16 1.3 0.16 11.9
2 16 4.2 0.17 4.1
3 16 7.1 0.17 2.4

Inter assay Precision was determined by assaying duplicates of three serum pools in 10 separate runs, using a standard curve constructed for each run.
Serum No. of Runs Mean
pg/ml
Standard Deviation Coefficient of Variation (%)
1 10 1.4 0.15 10.7
2 10 4.4 0.13 5.2
3 10 7.0 0.30 4.2


Sensitivity
The sensitivity was determined by calculating the mean plus 2SD of the standard zero point tested 20 times in the same run.
Serum Mean + 2SD (Sensitivity)
Zero Standard 0.05 pg/ml

Specificity
The cross reaction of the Triiodothyronine antibody to selected substances was evaluated by adding the interfering substance to a serum matrix at various concentrations. The cross-reactivity was calculated by deriving a ratio between dose of interfering substance to dose of Triiodothyronine needed to displace the same amount of tracer
Substance Cross reactivity Concentration (pg/ml)
I-Triiodothyronine 1.0000 -
I-Thyroxine <0.0002 10
Iodothyrosine <0.0001 10
Diiiodothyrosine <0.0001 10
Diiodothyronine <0.0001 10
Phenelbutazone <0.0001 10
Sodium Salicylate <0.0001 10
REFERENCESPederson, K.O, Scand. J. Clin. LAB Invest 34, 247 (1974)
Wild, D. Immunoassay Handbook, Stockton Press P339 (1994)
Wenzel K.W., Metabolism 30, 717, (1981)
Bhagat, C. et.al, Clin Chem 28, 1324. (1983)
Lundberg, P.R., et. Al, Clin Chem 28, 1241. (1982)
Melmed, S. et. Al, clin Endocrinol Metab 54, 300. (1982)
Lalloz M.R., et. al, clin Endocrinol 18, 11. (1983)
Citation banner

Thousands of laboratories across the world have published research that depended on the performance of antibodies from ADMEbio to advance their research.
Check out links to articles that cite our products in major peer-reviewed journals, organized by research category.

Thank you,
ADMEbio Team

Citation banner
Be the first to review product

ADMEbio welcomes feedback from our customers.

  • Be seen as an expert in the field of the products you are reviewing
  • Showcase your results (you can upload an image) to the antibody community. New product applications are welcome!
  • Share your experience to help other researchers make more informed decisions.
  • Easy process through our 5 star rating system and single login through your account.

Thank you,
ADMEbio Team